170135_inorganicventures_flipping
ICP-MS Basics
• Mass range 6 Li to 238 U
• Mass discrimination types
• High resolution (e.g., magnetic sector) • Lower resolution (quadrupole)
• Resolution (R) • R = m / Δ m
• m = mass being measured • Δ m = mass difference between two resolved peaks
Mass Resolution
• Resolution (R) • R = m /Δ m
• m = mass being measured • Δ m = mass difference between two resolved peaks
• Quadrupole typical R = ~400-600 • Magnetic Sector R > 10,000
Thomas, 2001
Made with FlippingBook - professional solution for displaying marketing and sales documents online